Acoustic Glossary


 

Time Weightings : Definitions, Terms, Units, Measurements..

Fast, Slow, Impulsive and Peak sound level descriptors are extensively used in acoustics.


Fast, Slow and Impulse

Most sound level meters have two exponential time weightings, F = Fast and S = Slow with time constants of 125 ms and 1000 ms respectively.

Some also have Impulse Time Weighting which is a quasi-peak detection characteristic with rapid rise time (35 ms) and a much slower 1.5 second decay.

F : Fast = 125 ms up and down,
S : Slow = 1 second up and down,
I : Impulse = 35 ms while the signal level is increasing or 1,500 ms while the signal level is decreasing.

Back to the days of analogue meters, Time Weightings were introduced to give the operator chance to 'follow' the rapid meter fluctuations by eye.


Peak : P-P : True Peak : Lpeak : Lpk : etc.,

Peak should not to be confused with Lmax which is usually measured with a Fast or Slow weighting.

To measure the True Peak values of impulsive sound levels a meter must be equipped with a Peak Detector.

The Peak Detector should responds in less than 100µs, according to the Sound Level Meter Standards, a typical response time, for a Class 1 meter is 40µs (40 microseconds).

Peak may be measured with a C or Flat Frequency Weighting. However the A curve, introduces it's own time constant which makes the measurement of True Peak impossible.

Peak Frequency-weighted Sound Pressure Level also known as Peak Sound Level is listed and defined below

Peak Hold is the peak detection process retaining the 'true' maximum value of a signal.


Peak Level IEC Definition, maximum instantaneous level of stated kind that occurs during a stated time interval

Peak Measurements are unambiguous for symmetric periodic waves like sine, square, etc., but ambiguous when the waveform is asymmetric

Peak Sound Level IEC Definition, greatest instantaneous value of a standard frequency weighted sound pressure level, within a stated time interval
Note : if frequency weighting is not specified, the A-frequency weighting is understood,

Also known as the Peak Frequency-weighted Sound Pressure Level


Peak Sound Pressure IEC Definition,: greatest absolute instantaneous sound pressure during a given time interval :


Peak-to-Peak : P-P is the amplitude difference between the most positive and most negative values in a time waveform.


Time Constant IEC Definition, time required for the amplitude of that component of a field quantity which decays exponentially with time to change by the factor 1/e = 0,367 9…


See also : Crest FactorExponential AveragingRMS - root mean square and the IEC Definition of Level